User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: MET-270-5 DATE: 01-Nov-94 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 16 NUMBER OF LINES: 207 CONFIGURATION: HP 33120A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD {Frequency Response} 1.002 IEEE [@33120A]*cls 1.003 IEEE [@33120A]outp:load inf 1.004 IEEE [@33120A]volt 20 1.005 IEEE [@33120A]freq 1000 1.006 IEEE [@33120A]freq:star 0.01 1.007 IEEE [@33120A]freq:stop 10000 1.008 IEEE [@33120A]trig:sour bus 1.009 IEEE [@33120A]swe:time 60 1.010 DISP Connect the HP 3458A DMM to the REPRO MON BNC 1.010 DISP on the rear of the recorder. Connect a 50ê 1.010 DISP termination to the HP 3458A DMM. 1.011 DISP Select CHANNEL 1 with the UP/DOWN arrow keys 1.011 DISP located at the upper right of the control panel. 1.012 OPBR Is there a Record PCB in this channel? 1.013 JMPF 2.003 1.014 DISP Connect HP 33120A Function Generator to CHANNEL 1. 1.015 HEAD {Channel 1} 1.016 CALL Freq_setup 1.017 JMPT 1.020 1.018 RSLT =PASS 1.019 JMP 2.001 1.020 EVAL Frequency Response failed, enter NO due to failure. 2.001 IEEE [@33120A]swe:stat off 2.002 HEAD { } 2.003 DISP Select CHANNEL 2 2.004 OPBR Is there a Record PCB in this channel? 2.005 JMPF 3.003 2.006 DISP Connect HP 33120A Function Generator to CHANNEL 2. 2.007 HEAD {Channel 2} 2.008 CALL Freq_setup 2.009 JMPT 2.012 2.010 RSLT =PASS 2.011 JMP 3.001 2.012 EVAL Frequency Response failed, enter NO due to failure. 3.001 IEEE [@33120A]swe:stat off 3.002 HEAD { } 3.003 DISP Select CHANNEL 3 3.004 OPBR Is there a Record PCB in this channel? 3.005 JMPF 4.003 3.006 DISP Connect HP 33120A Function Generator to CHANNEL 3. 3.007 HEAD {Channel 3} 3.008 CALL Freq_setup 3.009 JMPT 3.012 3.010 RSLT =PASS 3.011 JMP 4.001 3.012 EVAL Frequency Response failed, enter NO due to failure. 4.001 IEEE [@33120A]swe:stat off 4.002 HEAD { } 4.003 DISP Select CHANNEL 4 4.004 OPBR Is there a Record PCB in this channel? 4.005 JMPF 5.003 4.006 DISP Connect HP 33120A Function Generator to CHANNEL 4. 4.007 HEAD {Channel 4} 4.008 CALL Freq_setup 4.009 JMPT 4.012 4.010 RSLT =PASS 4.011 JMP 5.001 4.012 EVAL Frequency Response failed, enter NO due to failure. 5.001 IEEE [@33120A]swe:stat off 5.002 HEAD { } 5.003 DISP Select CHANNEL 5 5.004 OPBR Is there a Record PCB in this channel? 5.005 JMPF 6.003 5.006 DISP Connect HP 33120A Function Generator to CHANNEL 5. 5.007 HEAD {Channel 5} 5.008 CALL Freq_setup 5.009 JMPT 5.012 5.010 RSLT =PASS 5.011 JMP 6.001 5.012 EVAL Frequency Response failed, enter NO due to failure. 6.001 IEEE [@33120A]swe:stat off 6.002 HEAD { } 6.003 DISP Select CHANNEL 6 6.004 OPBR Is there a Record PCB in this channel? 6.005 JMPF 7.003 6.006 DISP Connect HP 33120A Function Generator to CHANNEL 6. 6.007 HEAD {Channel 6} 6.008 CALL Freq_setup 6.009 JMPT 6.012 6.010 RSLT =PASS 6.011 JMP 7.001 6.012 EVAL Frequency Response failed, enter NO due to failure. 7.001 IEEE [@33120A]swe:stat off 7.002 HEAD { } 7.003 DISP Select CHANNEL 7 7.004 OPBR Is there a Record PCB in this channel? 7.005 JMPF 8.003 7.006 DISP Connect HP 33120A Function Generator to CHANNEL 7. 7.007 HEAD {Channel 7} 7.008 CALL Freq_setup 7.009 JMPT 7.012 7.010 RSLT =PASS 7.011 JMP 8.001 7.012 EVAL Frequency Response failed, enter NO due to failure. 8.001 IEEE [@33120A]swe:stat off 8.002 HEAD { } 8.003 DISP Select CHANNEL 8 8.004 OPBR Is there a Record PCB in this channel? 8.005 JMPF 9.003 8.006 DISP Connect HP 33120A Function Generator to CHANNEL 8. 8.007 HEAD {Channel 8} 8.008 CALL Freq_setup 8.009 JMPT 8.012 8.010 RSLT =PASS 8.011 JMP 9.001 8.012 EVAL Frequency Response failed, enter NO due to failure. 9.001 IEEE [@33120A]swe:stat off 9.002 HEAD { } 9.003 DISP Select CHANNEL 9 9.004 OPBR Is there a Record PCB in this channel? 9.005 JMPF 10.003 9.006 DISP Connect HP 33120A Function Generator to CHANNEL 9. 9.007 HEAD {Channel 9} 9.008 CALL Freq_setup 9.009 JMPT 9.012 9.010 RSLT =PASS 9.011 JMP 10.001 9.012 EVAL Frequency Response failed, enter NO due to failure. 10.001 IEEE [@33120A]swe:stat off 10.002 HEAD { } 10.003 DISP Select CHANNEL 10 10.004 OPBR Is there a Record PCB in this channel? 10.005 JMPF 11.003 10.006 DISP Connect HP 33120A Function Generator to CHANNEL 10. 10.007 HEAD {Channel 10} 10.008 CALL Freq_setup 10.009 JMPT 10.012 10.010 RSLT =PASS 10.011 JMP 11.001 10.012 EVAL Frequency Response failed, enter NO due to failure. 11.001 IEEE [@33120A]swe:stat off 11.002 HEAD { } 11.003 DISP Select CHANNEL 11 11.004 OPBR Is there a Record PCB in this channel? 11.005 JMPF 12.003 11.006 DISP Connect HP 33120A Function Generator to CHANNEL 11. 11.007 HEAD {Channel 11} 11.008 CALL Freq_setup 11.009 JMPT 11.012 11.010 RSLT =PASS 11.011 JMP 12.001 11.012 EVAL Frequency Response failed, enter NO due to failure. 12.001 IEEE [@33120A]swe:stat off 12.002 HEAD { } 12.003 DISP Select CHANNEL 12 12.004 OPBR Is there a Record PCB in this channel? 12.005 JMPF 13.003 12.006 DISP Connect HP 33120A Function Generator to CHANNEL 12. 12.007 HEAD {Channel 12} 12.008 CALL Freq_setup 12.009 JMPT 12.012 12.010 RSLT =PASS 12.011 JMP 13.001 12.012 EVAL Frequency Response failed, enter NO due to failure. 13.001 IEEE [@33120A]swe:stat off 13.002 HEAD { } 13.003 DISP Select CHANNEL 13 13.004 OPBR Is there a Record PCB in this channel? 13.005 JMPF 14.003 13.006 DISP Connect HP 33120A Function Generator to CHANNEL 13. 13.007 HEAD {Channel 13} 13.008 CALL Freq_setup 13.009 JMPT 13.012 13.010 RSLT =PASS 13.011 JMP 14.001 13.012 EVAL Frequency Response failed, enter NO due to failure. 14.001 IEEE [@33120A]swe:stat off 14.002 HEAD { } 14.003 DISP Select CHANNEL 14 14.004 OPBR Is there a Record PCB in this channel? 14.005 JMPF 15.003 14.006 DISP Connect HP 33120A Function Generator to CHANNEL 14. 14.007 HEAD {Channel 14} 14.008 CALL Freq_setup 14.009 JMPT 14.012 14.010 RSLT =PASS 14.011 JMP 15.001 14.012 EVAL Frequency Response failed, enter NO due to failure. 15.001 IEEE [@33120A]swe:stat off 15.002 HEAD { } 15.003 DISP Select CHANNEL 15 15.004 OPBR Is there a Record PCB in this channel? 15.005 JMPF 16.003 15.006 DISP Connect HP 33120A Function Generator to CHANNEL 15. 15.007 HEAD {Channel 15} 15.008 CALL Freq_setup 15.009 JMPT 15.012 15.010 RSLT =PASS 15.011 JMP 16.001 15.012 EVAL Frequency Response failed, enter NO due to failure. 16.001 IEEE [@33120A]swe:stat off 16.002 HEAD { } 16.003 DISP Select CHANNEL 15 16.004 END